MB-200-CN Conductivity Transmitter  Home: Index >> Conductivity Transmitter

MB-200-CN Conductivity Transmitter

COND-8300 Conductivity Transmitter

Application area :
Widely used in waste water treatment, pure water, recycled water, boiler water systems And electronics, electroplating, printing and dyeing , chemical ,food, pharmaceutical and other process areas ,Large sewage treatment plants, industrial process monitoring applications such excellence.

 Instructions
 Color
    Features:

    l  96X96 microprocessor conductivity transmitter

    l  Waterproof & dustproof proof design, to fit wall/panel/pipe installation

    l  The coductivity,resistivity,salinity can be measured

    l  With backlight sensor, automatic and manual backlight setting function

    l  Automatic sensor clean function with programable time interval and hysteresis

    l  Large LCM display, it can show the operation status and symbols

    l  It can be connected to two poles and four poles conductivity electrode

    l  Provides digital filter function , can filter out interference value , to make the reading more stable

    l  Automatic calibration; manual/auto. temperature compensation

    l  With password and multi-key design、security protection enhancement.

    l Equipped RS-485 MODBUS RTU protocol, it can connectto the digital system directly (MB-200-CN serial only)

    l  Two programmable output (MB-200-CN serial only)

    l  TDS optional

    Parameters:


    Range/Resolution/Accuracy(±1Digit)  
    Conductivity  
    0.000~200.0mS/cm /0.001uS/cm /±1%±1Digit
    Specific resistance
    0.000~20.00MΩ.cm/0.01MΩ.cm/±1%±1Digit
    Salinity
    0.0ppt~70.0ppt /0.1ppt /±1%±1Digit
    Temp
    -30.0℃~130.0℃ /0.1℃ /±0.2%±1Digit;With manual correction function
    TDS
    0~19999ppm/0.001ppm/±1%±1Digit
    Electrode coefficient
    0.01,0.1,1,10.00 1/cm,0.0050~19.991/cm
    Temp. compensation  

    Automatic recognition with Pt1000/NTC30K or manual

    Temp. coefficient

    Conductivity:Linear(0.00%...20.00%)or  non-Linear ; Resistivity:non-Linear

    Hi/Lo Relay Output

    Two sets Hi/Lo programmable,N/OFF, 240VAC/ 0.5A Max

    Analog Output

    2 sets of programmable & isolate current DC 0/4-20mA with programable according to cond. /res. / sal./Temp. Max. load 500Ω .

    Wash Relay Output

    ON: 0...9999sec. / OFF: 0...999.9hours

    Specifications:


    AC supply
    100V~240VAC±10%,50/60Hz
    DC supply
    20~40VDC
    Temp
    0~50℃
    Dimensions
    96mm×96mm×148mm(H×W×D)
    Protection
    Correspond to IP65(NEMA 4)
    Installation
    wall/panel/pipe mounting
    Cut off Dimensions
    91mm×91mm (H×W)
    Weight
    0.4Kg