MB-300-PO-1 pH/ORP Transmitter
PH-620 Transmitter
Application area :
1,Industrial process ;2,Ferment process s Recycle & cleaner;3,Chemical process ;4,Drinking & surface water measurement;5,Water & waste water treatment 。
Features: |
l 144*144 microprocessor pH/ORP transmitter l Isolate current DC 0/4~20mA with programable according to pH/ORP、Temp (MB-300-PO-1 serial only) l Large LCD screen, with operation status and special special symbol display l Providing a sensor self-cleaning device contact,the program set out l Providing a solution grounding function, can eliminate sample charging interference l Waterproof & dustproof proof design, to fit wall/panel/pipe installation |
Parameters: | ||
Range | 0.00~14.00pH/-1999~+1999mV/-10.0~110.0℃ | |
Resolution | 0.01pH/1mV/0.1℃ | |
Accuracy | ±0.01pH±1Digit/±0.1%±1Digit/±0.2℃ Witch manual(±1Digit)correction function | |
Input Resistance | >10^12 Ω | |
Temperature Compensation | Automatic recognition with Pt1000/NTC30K or manual | |
Display | LCD display with sensitization sensorfor Auto/manual illumination function | |
Grounding Solution | a solution can eliminate interference Charged | |
Analog Output | 2 channel isolate current DC 0/4~20mA with programable according to pH/ORP、Temp. Max. load 500Ω | |
Wash Relay Output | ON: 0...9999sec. / OFF: 0...999.9hours |
Specifications: | ||
AC supply | 100V~240VAC±10%,50/60Hz | |
Temp | 0~50℃ | |
Dimensions | 144mm×144mm×115mm(H×W×D) | |
Protection | Correspond to IP65(NEMA 4) | |
Installation | wall/panel/pipe mounting | |
Cut off Dimensions | 138mm×138mm (H×W) | |
Weight | 0.8Kg |