MB-300-ION-1 Ammonia Concentration Transmitter  Home: Index >> ION Transmitter

MB-300-ION-1 Ammonia Concentration Transmitter

ION-7110 Ammonia Concentration Transmitter

Application area :
The ion selective sensor is installed directly in the activated sludge tank of the sewage treatment plant,Without any sample pretreatment or sample transfer.The whole measuring system includes sensors and transmitters:Sensor containing electrode,Transmitter includes display and operating unit.Sensor for monitoring the concentration of ammonia nitrogen: sSludge tank sThe primary settling tank outlet.

 Instructions
 Color
    Features:

    l  Without sample pretreatment, direct ammonia nitrogen measurement

    l  Potassium ion measurement is optional, can also compensate for the high concentration of interference

    l  pH measurement and compensation

    l  Without reagents, the use of low cost

    l  Range:0.1-1000 mg/l NH4-N

    l  Direct installation in the pool, without liquid collector or sample transfer pump

    l  Compressed air cleaning optional, low maintenance cost

    l Every 6 months to replace the coated cap, effectively extend the life of the electrode

    Parameters:


    The input measurement variables depend on the type of sensor

    ammonia nitrogen
    potassium ion

    NH4-N,NH4+ [mg/l] 0.1~1000 mg/l

    K+ [mg/l]  1~1000 mg/l(可选)

    response timet90
    ammonia nitrogen:< 2 min. potassium ion:< 2 min.
    Accuracy
    measured values of + 5% + 0.2 mg/l
    Repeatabilit
    display value of + 3%
    Correction mode
    Two or three-point calibration, offset correction (providing three zones Section optional)
    input resistance
    >10^12Ω
    Hi/Lo Relay Output
    2 sets Hi/Lo programmable,N/OFF, 240VAC/ 0.5A Max
    Analog Output
    2 sets of programmable & isolate current DC 0/4~20mA with programable according to DO、Temp. Max. load 500Ω
    Environment condition:

    Ambient temperature range
    -20 ... 50 °C (-4 ... 120 °F)
    Storage temperature
    2 ... 40 °C (36 ... 100 °F)
    Protection class
    IP 65
    Process condition
    Process temperature range
    2 ... 40 (36 ... 100°F)
    Maximum allowable pressure of process pressure
    400 mbar (160 inH2O)
    pH value of medium:
    ammonia nitrogen
    pH 5...8.3 (No pH compensation)  pH 5 ...10(with pH compensation)
    potassium ion
    pH 2 ~12